Quick Details
- Voltage range:200 mV, 400 mV, 2 V, 4 V, 20 V, 100 V, 200 V, 500 V, 1000 V;
- Current range:40mA, 0.2A ,0.5A, 2A, 5A, 10A, 20A;
- Clamp ampere ammeter range:200A, 600A, 1000A;
- Frequency Output Range:45-65Hz
- Harmonic Wave Times:2-31
- Resistance box Range:0-4MΩ
- Weight (kg):12
- Processing Time:7 days
- Port:Tianjin
- Supply Ability: 5000 Sets Per Month
- Brand Name:GFUVE
- Model Number:GF6018A
- Place of Origin:China
- Brand Name:GFUVE
GFUVE Laboratory Multifunction Electrical Tester GF6018A Multimeter Calibrator
Introduction of Power calibration products GF6018A Clamp Meter
Calibrator with excellent working performance ,CE,ISO9001:2008,IEC approved
Photos
Product Description
FEATURES
It can be semi-automatic or manual inspection of analog multi
meter, clamp type current meter, all kinds of electric meter, (voltmeter, ammeter, frequency meters, resistance meter) basic error
The equipment can be generated with 2-31 harmonic wave,
harmonic number, order, amplitude and harmonic phase of fundamental wave can be programmed
Power amplifier working frequency range is 40Hz-1KHZ, has a
good linear. Current amplifier is constant current source; voltage amplifier is constant voltage source. Because of light weight, this device is more suitable for field use
With RS-232 interface, this device is controlled by PC
through the PC software (optional), can be automatic or manual inspection, and the results are processing and management
Has a non-volatile memory, it can store 500 pieces of the raw
data of tested meter, for access and upload
PARAMETERS
Electrical parameters
Power supply
Single phase AC 220V±10% or 110V±10%, 50/60HzAC
AC Voltage output
range
200mV, 400mV, 2000mV, 4000mV,
20V, 100V, 200V, 500V, 1000V
Range 20V-1000V
Accuracy
±(0.03%RD+0.02%FS)
Stability
≤0.01% FS /60s
Range 200mV-4000mV
Accuracy
±(0.3%RD+0.2%FS)
Stability
≤0.04% FS /60s
Load capacity
20VA
Waveform distortion
≤0.3%
Adjustable Range
0-120% FS (1000V not included)
Fineness
5×10-5
AC Current Output
Range
2mA, 4mA, 20mA, 40mA, 0.2A, 0.5A, 2A, 5A, 10A, 20A
Range 0.2A-20A
Accuracy
±(0.03%RD+0.02%FS)
Stability
≤0.01% FS /60s
Range 2mA-40mA
Accuracy
±(0.3%RD+0.2%FS)
Stability
≤0.04% FS /60s
Load capacity
20VA
Waveform distortion
≤0.3%
Adjustable Range
0-120% FS
Fineness
5×10-5
Frequency Output
Range
45-65Hz
Fineness
0.001Hz
Accuracy
0.01Hz
Harmonic Wave
Times
2-31
Extent
0-20%
Each harmonic phase fineness
0.01·N (N is harmonic order)
DC voltage output
Range
200mV, 400mV, 2000mV, 4000mV,
20V, 40V, 100V, 200V, 400V, 1000V
Load capacity
20W
Adjustable range
0-110% FS (1000V not included)
Adjustable fineness
5×10-5
Stability
≤0.01% FS /60s (peak-peak value)
Accuracy
±(0.03%RD+0.02%FS)
Ripple wave
≤1% FS
DC current output
Range
20uA, 40uA, 200uA, 400uA, 2mA, 4mA,
20mA, 40mA, 0.2A, 1A, 2.5A, 5A, 10A, 20A
Load capacity
20W
Adjustable Range
0-110% FS (1000V not included)
Adjustable Fineness
5×10-5
Stability
≤0.01% FS /60s (peak-peak value)
Accuracy
±(0.06%RD+0.04%FS)
Ripple wave
≤1% FS
Resistance box
Range
0-200Ω, 200-400Ω, 0-2kΩ, 2kΩ-4kΩ, 0-20kΩ, 20kΩ-40kΩ,
0-200kΩ, 200kΩ-400kΩ, 0-2MΩ, 2MΩ-4MΩ
Fineness
0.02%
Range 0-2kΩ, 0-20kΩ, 0-200kΩ
Range 0-2kΩ
Adjustable Range
0-100%FS
Accuracy
±(0.12%RD+0.08%FS)
Range 0-200Ω,0-2MΩ
Adjustable Range
0-100% FS
Accuracy
±(0.3%RD+0.2%FS)
Range 2kΩ-4kΩ, 20kΩ-40kΩ, 200kΩ-400kΩ
Range 2kΩ-4kΩ
Adjustable Range
5%-100%FS
Accuracy
±(0.12%RD+0.08%FS)
Range 200-400Ω,
2MΩ-4MΩ
Adjustable Range
5%-100% FS
Accuracy
±(0.3%RD+0.2%FS)
Clamp meter range
200A
Accuracy
±0.2%
600A
Accuracy
±0.5%
1000A
Accuracy
±1%
Dot frequency square wave output
Range
1kHz, 5kHz, 10kHz, 50kHz, 100kHz, 200kHz, 1000kHz
Square wave amplitude
4.5V
Setting accuracy
5×10-5
Mechanical parameters
Dimension (L×W×H) (mm)
440x360x160
Weight (kg)
12
Environmental conditions
Operating temperature
20°C±10°C
Relative humidity
≤85%RH