Zwin-YCB06 β-Ray air quality monitoring system adopts the principle of β-ray absorption to measure the particle concentration. The result is accurate and reliable.

The monitoring and data management are operated by the microcomputer system , which is able to inspect continuously.

Application

Mainly used in air quality monitoring networks, mobile monitoring stations, environmental research, industrial and mining enterprises, research institutes and other fields.

Â

Features

· beta ray absorption method to measure TSP, PM10, PM2.5 and PM1

·5.7” LED display

·Data storage: storage capacity can reach to 1,000,000. On board storage for over 20 years of data

· Keyboard input: easy input, man-machine dialogue mode

· One channel for sampling and analysis in case of any error caused by paper feed.

· Automatic fault diagnosis and alarms, also capable of fixing errors remotely

· Multi interface output: analog output, current output, relay output

Specifications

Concentration Range

Optional:(0-1000)µg/m3 (0-10000)µg/m3

accuracy

±2%

Full-scale value

1.5mg/cm2

lowest detectable limit

≤2µg/m3

Temperature before measuring

-30℃~+50℃,error of indication ±0.5℃

Pressure before measuring

60kPa~110kPa,error of indication ±2.5%

reproducibility

≤2%

parallelism

≤±15%

sampling deviation

Â

±2%( 16.7L/min as constant value)

Timming error

24h<10s

noise

≤65dB

Operating Temperature

temperature:20℃~50℃

humidity:≯80%

pressure:86kPa~106kPa

Power supply:AC220V±22V, frequency 50Hz±1Hz

Detector

PM10

PM2.5

Geiger counter tube

tomultiplier

Data display period

60min

acquisition cycle

60min

Calibration

Standard diaagm(single or multi point)

Interface

Serial communication, analog

operation

button

Restart

Automatically restart after power restored

Fault Alarm

Abnormal data, break warning

Data sotrage

On-board storage,easy to quarry and drive

Temperature range

≤40℃

Humidity range

35%RH±5%

Control Voltage

+15V、+5V

Input voltage

220V±10%、50Hz

Power

≤150W

Sensor precision

Temp ±1℃;

Humidity ±3%RH