Model:PC100EParameters:SPO2+NIBP+ETCO2Features:1. The Smallest and exquisite handheld vital sign monitor in the world2. Accurate measurement in 0.05% low perfusion index,the world leading production in the movement environment3. Provide the parameter of the PI(perfusion index 0.05%-20%)4. Accurate measurement in all kind of extreme environment measurement such as neonatal and shock patients5. NIBP adopt SAWAA(Self-Adaptive-Waveform-Amplitude-Adjust)algorithm and SAFD(Self-Adaptive-Fast-Deflation)algorithm6. Easy operation by 3.5"color TFT screen and Up to 100users ID switch7. Multiple vital signs measurement parameters for collocation optional l NIBP,SPO2,TEMP l SPO2,TEMP l NIBP,TEMP l SPO2,ETCO2 l SPO2,NIBP,ETCO28. Two measurements mode:Spot,Monitoring9. Data storage capabilities,support for historical data view and trend chart back to see l In monitoring mode,can continuously record 48 hours of measurement data l In spot mode,100users ID,each user can store 200sets data10. Support audio and visual alarm,support to set alarm uper and lower limit values11. Lithium ion battery power supply12. Optional charging dockPerfomance Specifications:
Display | 3.5" color TFT | Resolution | 320*480 |
Indicator | Alarm indicator,Power indicator | Trace | 1 plethysmogram waveform |
Alarm | Probe off,low power | modes | visual and audio |
Application | Neonatal,pediatric and adult patients |
Parameters Performance:NIBP
Adult | PED | NEO | |
SYS | 40~270mmHg | 40~200mmHg | 40~135mmHg |
MEAN | 20~230mmHg | 20~165mmHg | 20~105mmHg |
DIA | 10~210mmHg | 10~150mmHg | 10~95mmHg |
Pulse Rate Range:40~240bpm | |||
NIBP measurement accuracy | ±3mmHg | ||
Pulse rate measurement accuracy | â¤2%(â¥100bpm)or 2beats/min(<100BPM) |
Independent software protection pressure | ||
Adult:300mmHg | PED:240mmHg | NEO:150mmHg |
Independent hardware protection pressure | ||
Adult:320-330mmHg | PED:265-275mmHg | NEO:160-165mmHg |
Oxygen Saturation(%SPO2)
Measurement Range | |||
Measure Range | SPO2:0-100% | PR:0-254BPM | Perfusion index:0.05%-20% |
Accuracy Range | SPO2 saturation:70-100% | Pulse Rate:20-254BPM | Perfusion Index:0.025-20% |
Measurement Accuracy | |||
SPO2 | |||
Adult(including pediatric):±2digit(70-100%) not defined(<70%) | Neonate:±3digits not defined(<70%) | On motion condition :±3digits | |
Pulse Rate | |||
Adult(including pediatric):±3digits | Neonate:±3digits | On motion condition::±3digits | |
Perfusion Index:0.05%-20% | |||
Resolution | |||
SPO2 saturation:1% | Pulse Rate:1BPM |
CO2Measurement range:ETCO2:0-150mmHg(0-19.7%)(0-20kPa) RR:3-150BPMAccuracy range:
ETCO2 Range | Accuracy |
0-40mmHg | ±2mmHg |
41-70mmHg | ±5% of reading |
71-100mmHg | ±8% of reading |
101-150mmHg | ±10% of reading |
RR Accuracy:±1BPMPaymentShipping