Join today and be a part of the fastest growing B2B Network Join Now

CE approved multi-function ICU medical equipment anesthesia machine in China 6200A anesthesia workstation with ventilator

  • Origin: China
  • Supply Type: in stock
  • Processing Time: within 3-5 working days
  • Min Order: 1

Quick Details

Ventilation modes: P-CMV, V-CMV, P-SIMV, V-SIMV, PSV, SIGH, MANUAL
Warranty 12 months
Certificate CE&ISO
Tidal volume(Vt) 0, 20 mL ~ 1500 mL
Frequence(Freq) 1bpm ~ 100 bpm
I/E 4:1~1:8
PEEP 0cmH2O ~ 30 cmH2O
Display 12.1″ TFT LCD touch screen

Supplier Info.

  • Employees Total 11-50
  • Annual Revenue US$1 Million - US$2.5 Million

Features

Â

  • Electronic flowmeter

High precision flowmeter, instantly know the fresh gas flow to patient.

O2 and N2O linkage device ensure O2 concentration no less than 25%.

Â

  • 12.1″ TFT LCD touch screen

Displays the Ventilation parameters, Alarm information and Oscillogram.

High sensitivity touch screen ensures accurate and easy operation. Alternate button for dual control.

Technical specifications

Ventilation modes:P-CMV, V-CMV, P-SIMV, V-SIMV, PSV, SIGH, MANUAL
Ventilator parameter rangesÂ
Flow meterO2(0.1-10L/min)
 N2O(0.1-10L/min)
 AIR(0.1-10L/min)
Rapid oxygen supply35L/min-75L/min
Tidal volume(Vt)0, 20 mLÂ ~Â 1500 mL
Frequence(Freq)1bpm ~ 100 bpm
I/E4:1~1:8
PEEP0cmH2OÂ ~Â 30 cmH2O
Pressure triggering sensitivity(PTr)-20 cmH2OÂ ~Â 0 cmH2O(Based on PEEP)
Flow trigger sensitivity (FTr)0.5 L/min ~ 30 L/min

Pressure control (PC)

5 cmH2OÂ ~Â 60 cmH2O
SIGH0(off)Â 1/100Â ~Â 5/100
Apnea VentilationOFF,5 s ~ 60 s
Pressure Limit20 cmH2OÂ ~Â 100 cmH2O
Monitored ParametersÂ
Frequence(Freq)0 /min ~ 100 /min
Tidal volume(Vt)0 mLÂ ~Â 2000 mL
MV0 L/min ~ 100 L/min
Oxygen concentration15 %Â ~Â 100 %

Â

Gracal display:Â
P-T(pressure – time)Â
F-T(flow – time)Â
P-V loop (pressure – volume loop)Â
V-T(volume – time )Â
Â

Alarm and protectionÂ

The AC power failure alarmPower failure or no connection
Internal battery backup low voltage alarm<11.3±0.3V
No tidal volume≤5Ml within 6s
High oxygen concentration alarm

Low oxygen concentration alarm

19%-100%

18%-99%

Suffocation warning5s-60s no spontaneous ventilation
The maximum limited pressure <12.5 kPa
Fan error

Oxygen deficit

Â

Show on screen

Show on screen

Working conditionsÂ
Gas sourceO2,N2O,AIR
Pressure280kPa-600kPa
Voltage100-240V
Power frequency50/60Hz
Input power80VA
VaporizerHalothane, Enflurane, Isoflurane, Sevoflurane
Premium Services
Tell us your Requirement
Girl Right
Cross Popup
Arrow 2

I Am :

Signup today to claim your Discount. Get Started before it's too late!

Arrow 1