GB 2626-2006 Respiratory protective equipment Non-powered air-purifying particle respirator
GB 2890-2009 Respiratory protection-Non-powered air-purifying respirators
Purpose:Take the exhalation valve installation on the suitable fixture , then will take the fixture inner change to negative pressure state, monitor the whole fixture device recover to normal pressure time , to determine exhalation valveâs air tightness
Technical parameter:
1.2 Flow meter range:0 ï½ 800mL/min
1.3 Micro manometer range:0 ï½ 2000Pa
1.4 Time accurate:0.1s
1.5 exhale valve with constant cavity should tightness , with horizontal vertical, valve sheet shouldnât suffer the force deformation
2.1 pressure gauge range:0 ï½ 0.1MPa
2.2 Airway inner pressure value within 15s havenât change
2.3 Airway length 50~100cm
3.1 Constant cavity size:150±10mL
3.2 Micro manometer range:0 ï½ 2000 Pa
3.3 Flow meter range: ï¼ 0 ï½ 1000mL/min
3.4 Flow meter accurate:3%
3.5 Timer accurate:0.1s
3.6 Sucker pump speed rate:2L/min
3.7 Can satisfy 6.14 section air tightness testing
PLC control