Join today and be a part of the fastest growing B2B Network Join Now

Glycosylated Hemoglobin (HbA1c) Test Kit Poct Diagnostic Reagent (Fluorescent Immunoassy)

  • Origin: China
  • Supply Type: oem service
  • Processing Time: 5
  • Min Order: 100

Supplier Info.

  • Employees Total 101-200
  • Annual Revenue US$10 Million - US$50 Million
Product NameÂ
Glycosylated Hemoglobin(HbA1c)Test Kit (Immunofluorescence Chromatogra)Â
PurposeÂ
To detect the content of HbA1c in human whole blood in vitro quantitatively.For Professional Use only.Â
Principles of DetectionÂ
The HbA1c Test is a fluorescent lateral flow immunoassay. The test consists of: 1) a conjugate pad containing HbA1c antibody conjugated with fluorescent microsres and a control antibody (goat anti-chicken IgY antibody) conjugated with fluorescence microsres, 2) a nitrolose membrane strip containing a test ine) and a control ine). The T re-coated with another HbA1c antibody, and the C re-coated with a control c; antibody (chicken IgY antibody).Â
HbA1c concentration in specimen is positively correlated with the fluorescent density of the microsres, after the measure and calculation of the quantitative fluorescence immunoassay analyzer, then the analyzer would calculate the concentration of HbA1c according to the preset calibration curve in the machine and display the result, detection unit is %units. C ld appear in any cases, which indicates the reaction system is effective.Â
Storage conditions and shelf lifeÂ
Store in a sealed aluminum foil bag at 2~30°C. After unpacking the aluminum foil bag under temperature 20~30 ºC , humidity 35~65%, please use it immediately.Â
Shelf life 18 months.Â
Applicable instrumentÂ
Avid fluorescent immunoanalyzer model: AI-IF 600Â
Manufacturer& After-Sales Company: Shenzhen Aivd Biotechnology Co., LTD.Â
Address:C501,Building B5,China Merchants Guangming Science Park, #3009 Guanguang Road, Guangming District, Shenzhen,Guangdong Province, ChinaÂ
 Â
Specimen collection and preparationÂ
1. Human whole blood samples can be used for this test.Â
2. Specimen collection:Collect blood specimen into a collection tube containing EDTA , evenly mixed before use. Do not use hemolyzed blood for testing.Â
Whole blood specimens should be stored at 2-8°C for no more than 4 hours if not test immediately.Â
3. Do not test patients' samples with severe hemolysis, lipemia and jaundice, all specimens should be treated as infectant.Â
4. Restore the specimen to 20-30°C before the detection, cryopreserved specimen should be completely thawed, rewarmed and evenly mixed before use.Â
Test procedureÂ
Read the complete instruction before the testÂ
1. Bring the specimen and test components to room temperature (recommended temperature is about 20-30 ºC ) to test. Open the aluminum foil bag containing the test kit after it get back to the room temperature. Keep the test kit in a dry place to prevent spoilage from dampness. Before unpacking, check whether the aluminum foil bag is in good condition. Do not use it if it is damaged. The test card should be used within 1 hour after unpack.Â
2. Open the instrument as its instruction and Insect the ID card to get the curve.Â
3. Collect 10μl of sample(whole blood), add to the sample diluent (mix thoroughly for 30 sec with vortex mixer) collect 80μl of a sample mixture and load it into the sample vertically and slowly.Â
4. Leave the sample-loaded cartridge at room temperature (recommended temperature is about 20-30ºC) for 15 minutes. After time is over, immediately insert it into the cartridge holder of the instrument, press "Test" button and read the results.Â
5. Treat the tested card as potential biological hazardÂ
6. Quality Control:Â
1) As an effectiveness indicator, CÂ ld appear in any cases.Â
2) Aivd's internal control materials or other approved applicable control materials can be used to control the quality of the product, the test results should be within the specified quality control range.Â
Note: Perform the test under the condition of temperature 20~30ºC, humidity 35~65%.Â
Reference rangeÂ
Positive: reading value ≥6%, a positive result.Â
Negative: reading value <6%, a negative result.Â
Reading value 4%-6%, normal blood glucose.Â
Reading value 6%-8%, controllable blood glucose.Â
Reading value >Â 8%, poorly controlled blood glucose.Â
The reference value is verified with reference to similar products that have been marketed. Due to ethnic and geogracal differences, each laboratory can establish its own reference interval according to the actual situation.Â
Explanation of the resultÂ
1. If the sample is cloudy, it will affect the flow rate of the sample, resulting in prolonged or undetectable test time, which may lead to incorrect test results, please centrifuge and discard the precipitate before use.Â
2. The accuracy of the spiked volume will directly affect the accuracy of the test results.Â
3. Please check the integrity and expiration date of the kit package before use, then open the package, and when stored at low temperature, it should be restored to 20~30ºC before opening the package for use. Direct use at low temperature will affect the test results.
4. Whether the patient has inflammatory bacterial infection should be judged by the sician in combination with clinical features and symptoms, as well as other diagnostic methods of the patient.Â
Limit of the Test MethodÂ
1. The test results can only be used as an aid to sicians or other diagnoses, and need to be combined with other clinical and laboratory data; if the test results do not match the clinical assessment, further examination is required.Â
2. Bilirubin ≤0.5mg/mL, triglycerides ≤10mg/mL, hemoglobin ≤5mg/mL, within ±10% deviation to the test results.Â
3. The HOOK effect will not occur when the HbA1c concentration content in the sample is ≤20%.Â
4. If the sample test result shows more than 14%, it is recommended to dilute the sample with sa maximum dilution should not exceed 1:5), and the concentration value of the test obtained after its dilution, then multiply the dilution times to calculate the sample concentration value.Â
Performance CharacteristicsÂ
1. ngeÂ
In the range of 4%-14%, the r value should be ≥ 0.990.Â
2. AccuracyÂ
Detected with the enterprise internal control accuracy reference (C1, C2), the test results should be within the following ranges.Â
(1) Detection with the enterprise internal control accuracy reference C1, the instrument measurement range is 6.4% ±15% at 15 minutes.Â
(2) Detection with the enterprise internal control accuracy reference C2, the instrument measurement range is 8.4% ±15% at 15 minutes..Â
3. Intra-batch precisionÂ
Using the enterprise internal control precision reference (C9, C10) each parallel test 10 times, at 15 minutes instrumentation, the concentration value coefficient of variation CV≤15%.Â
4.Inter-batch precisionÂ
For three batches of reagent kits, 10 times of parallel testing with each enterprise internal control precision reference (C9, C10), and the instrument measurement at 15 minutes, the concentration coefficient of variation CV≤15%.Â
5.Detection limitÂ
Detection with the enterprise internal control of the detection limit reference L1, instrumentation at 15 minutes, the results should be ≤4%.Â
6.SpecificityÂ
Detection with the enterprise internal control specificity reference (C3, C4, C5), the instrument measurement value at 15 minutes should be less than 6%.
Â
NotesÂ
1. This kit is only for in vitro diagnosis, one-time use, please do not reuse.Â
2. Kits should be treated as containing infectious material.Â
3. Please check the integrity of the kit packaging and expiration date before use.Â
4. Please read the instructions for use of this reagent and instrument in detail before all operations.Â
5. Please operate in strict accordance with the requirements of the instructions. After the test starts, it cannot be stopped halfway. If it is stopped halfway, the test cannot be resumed. If you need to retest, you must update the reagents and retest.Â
6. Each batch of reagents has corresponding parameters in the supporting instrument, and the manufacturer updates the parameters in the instrument regularly. If the new batch of reagents is not recognized by the instrument, please the manufacturer to update the parameters in time.Â
7. The test card should not be used for more than 1Â hour after opening.Â
8. Reagents with different batch cannot be mixed, and ID cards and test cards cannot be used with different batch .Â
9. Excessive experimental temperature should be avoided, and the test card stored at low temperature needs to be restored to room temperature before opening to avoid moisture absorption.
Premium Services
Need Free Quotes
Girl Right
Cross Popup
Arrow 2

I Am :

Signup today to claim your Discount. Get Started before it's too late!

Arrow 1
We use cookies to ensure that we give you the best experience on our website. If you continue to use this site we will assume that you are happy with it. Ok