Weight | 0.6kg |
DDG-2080 Industrial Conductivity Meter
Â
DDG-2080 Industrial onan>Conductivity Meter is our companyâs latest top-class instrument equipped with a microprocessor and complete English menu. It has English display and English menu operation pattern; it is characterized by its intelligent operation, multi-function and high measuring performance and its strong environmental adaptability. It can be widely used for continuous monitoring of the conductivities of solutions used in the following fields: thermal power plant, chemical fertilizer, chemical industry, metallurgy, environmental protection, rmacy manufacturing, biochemistry, foodstuff and running water.  Â
Â
Features
It has a complete English display and English menu operation, various parameters can be displayed at the same |
time: conductivity, output current, temperature, time and status. Bitmap type liquid crystal display module |
with high resolution is adopted. All the data, status and operation prompts are displayed in English. |
Â
Â
Technical Indexes
1.Conductivity measuring range: 0.01~20μS/cm    (Electrode: K=0.01) |
                           0.1~200μS/cm   (Electrode: K=0.1) |
                           1.0~2000μS/cm  (Electrode: K=1.0) |
                           10~20000μS/cm (Electrode: K=10.0) |
                           30~600.0mS/cm (Electrode: K=30.0) |
2. Intrinsic error of the electronic unit: conductivity: ±0.5ï¼ FS, temperature: ±0.30â |
3. Range of automatic temperature compensation: 0~99.9â, with 25â as the reference temperature;0~150â (optional) |
4. Water sample tested: 0~99.9â, 0.6MPa |
5. Intrinsic error of the instrument: conductivity: ±1.0ï¼ FS, temperature: ±0.5â |
6. Automatic temperature compensation error of the electronic unit: ±0.5ï¼ FS |
7. Repeatability error of the electronic unit: ±0.2ï¼ FS±1 Unit |
8. Stability of the electronic unit: ±0.2ï¼ FS±1 unit/24h |
9. Isolated current output: 0~10mA ( load<1.5kΩ) |
                     4ï½20mA (load<750Ω) Can double the current output. |
10. Output current error: â¤Â±lï¼ FS |
11. Error of electronic unit caused by ambient temperature: â¤Â±0.5ï¼ FS |
12. Error of the electronic unit caused by the supply voltage: â¤Â±0.3ï¼ FS |
13. Alarm relay: AC 220V, 3A |
14. Communication interface: RS485Â (optional) |
15. Power supply: AC 220V±22V, 50Hz±1Hz, DC 24V(optional) |
16. Protection grade: IP65 |
17. Clock accuracy: ±1 minute/month |
18. Data storage capacity: l month (1 point/5 minutes) |
19. Saving time of data under a continuous power-failure condition: 10 years |
20. Overall dimension: 146 (length) x 146 (width) x 108 (depth) mm; dimension of the hole: 138 x 138mm |
21. Working conditions: ambient temperature: 0~60â; relative humidity <85ï¼ |
22. Weight: 0.6kg |
23. The conductivity electrodes with the following five constants are usable: K=0.01, 0.1, 1.0, 10.0, and 30.0. Â |