Winner 2008 laser particle size analyzer

  • Origin: China
  • Supply Type: in stock
  • Min Order: 1

Supplier Info.

  • Employees Total 51-100
  • Annual Revenue US$2.5 Million - US$5 Million

Winner2008A full automatic wet laser particle size analyzer adopt MIE scattering principle, measure size is from 0.01μm to 2000μm, Which offer reliable and repeatable particle size analysis for a diverse range of applications.It use dual-beam& multiple spectral detection systems and side light scatter test technology to significantly improve precision and performance of test, on behalf of the domestic advanced level in the field.

1)Winner Patent -Optical path designConverging light Fourier transform light patented technology, make scattering light not be restricted to lens aperture. Double spectrum design make its precision achieves from 0.01um to 1200μm,and Dual-laser orthogonal light make use of the semiconductor auxiliary laser extend the test angle from 45 degree to 135 degree, ensure receive all the angles of r>2) Expand Size Range measurement range significantly expanded to 0.01um, particularly suitable for small samples of the user, the sub micron particle test results reached the international level 3)Full built-in dispersion systemThe ultrasonic dispersion, mechanical stirring and circulation channel reasonable integrated in the interior of the instruments, ensure the uniform dispersion and distribution of particles in the testing process, effectively avoid uneven distribution, large particle deposition nomenon caused by long pipe utside dispersion system , to ensure the accuracy of the test results. 4)Intelligent Operation modeWith intelligent automatic mode of operation, to achieve a key test, as long as according to the prompt addition of sample, click the "test", all processes including water-supply, dispersion, circulation, testing, cleaning, data record, data analysis, save and print are automatically completed, not only reduce the testing workload, but also eliminate the interference of human factors, to further improve the accuracy and authenticity of testing results. 5)Automatic Optical path alignment system The precision of four hybrid stepping motor in the automatic system of optical components, micro precision of 0.1um, the instrument of light path is always at its best to eliminate manually on the light path and the troubles and difficulties but also enhance the accuracy and stability of test results. 6)Authoritative calibration method In the entire test range using the national standard materials for calibration, only a year calibration once the instrument can. Calibration method is simple and quick. 7) Unconstrained free fitting technique Winner original unconstrained free fitting technology, particle size analysis is not affected by any functional constraints, can be a true reflection of the distribution of particles.
Model Name
Winner2008A
Winner2008B
Standard
ISO13320-1:1999, GB/T19077.1-2008, Q/JWN001-2009
Principle
MIE scattering principle
Measuring Range
0.01μm-2000μm
0.01μm-1200μm
Channels /div>
127
120(different to detectors & optical light path)
Accuracy error
≤0.5% (Deviation of D50 on national standard sample)
Repeatability error
≤0.5% (Deviation of D50 on national standard sample )
Light source
High performance He-Ne Laser (λ= 632.8nm, P>2MW)Lifetime>25000hours
DispersionMethod
Ultrasonic
Frequency:40KHz Power:35W, Time: ≥1S ,Anti-dry function
Stir
Revolutions Speed: 0-3000RPM (Adjustable)
Circulate
Rated Flow:8L/min Rated Power:10W
Sample Pool
Volume:350mL
Micro-Sample Pool
Volume: 10mL ( Available)
Operation Mode
Full automatic/ manual control, freely choose
Resolution
Free distribution truly reflect particle size distribution
Optical bench alignment system
Japan Canon lens, Full automatic,precision is up to 0.1um
Software function
Analysis mode
Free Distribution, R-R Distribution, Logarithm Normal Distribution, Mesh classification etc.
Statistic Method
Volume Distribution, Quantity Distribution
Statistic Comparison
Several Testing Results of samplesDifferent batches of samples testing result,Samples before and after processing,Test result of samples in different time.
User-defined Analysis
Figure out percentage according to the particle sizeFigure out particle size according to the percentageFigure out percentage according to the particle size rangeMeet demands of representation of particle test in different industries
Test Report
Word, Excel,to( Bmp), Text etc
Multiple-language Support
Multiple language Support
Intelligent operation
Automatically control water inflow, dispersion,test and analysis.Better Repeatability after remove human-factor
Testing speed
<2min/time
Outer dimension
L104cm*W44cm*H54cm
Net weight
58Kg