+393512941528 | |
+447452273447 | |
Use | Hearing aid |
Warranty | 2-years |
+4
Features:
* Fast and easy ABR + OAE hearing screening
* Powerful ABR detection algorithm for testing within
seconds
* Truly intuitive user interface
* On-screen guidance; i.e. probe fitting feedback, noise
monitoring
* Small and lightweight
* Large color touchscreen for comfortable navigation
* Very short boot-up time and long battery life
* Cradle for ing
* HearSIM patient database and device manager
Technical Data:
easyScreen
Dimensions / Weight 163 x 85 x 21 mm / 265 g
Display 4.3â color touch screen
Languages English, German, Spanish, French, Italian,
Dutch, Turkish, Chinese, Japanese, Korean,
Romanian, Norwegian
Storage max. 250 patients with 50 tests each
PC interface micro USB
Battery Rechargeable Li-ion
Cradle
Charging ctive charging)
Power supply 5 V DC (micro USB)
Dimensions / Weight 142 x 140 x 62 mm / 270 g
ABR
Stimuli CE-Chirp®
Stimulus rate ~90 /s
Test mode monaural or binaural
Level 35 dB nHL
Impedance test Automatic pre-test
Test duration 10 s to 180 s
Artifact rejection level 100 µV
Insert
Transducer Insert IP30 with eartips or EarCupsâ¢
Preamplifier
One Channel 3 electrode cables (black, yellow and white)
Gain 72 dB
CMR Ratio >110 dB at 100 Hz
Dimensions / Weight 85 x 50 x 25 mm / 85 g
TEOAE*
Stimulus type Click
Level 83 dB SPL (in-ear calibration)
Analysis bands 1.4, 2, 2.8 and 4 kHz (center frequencies)
Pass criteria 3 out of 4 bands, min 4 dB SNR,
min -5 dB SPL OAE level
Test duration 4 s to 60 s
DPOAE*
Stimulus type Pairs of pure tones
Level (L1/L2) 65/55 dB SPL
F2 Frequencies 2, 3, 4 and 5 kHz
F2/f1 ratio 1.22
Pass criteria 3 out of 4 bands, min 6 dB SNR,
min -5 dB SPL OAE level
Test duration 4 s to 60 s
OAE Probe
Cable length 120 cm
Standards
IEC 60645-6 2009 Type 2, IEC 60645-7 2009 Type 2,
IEC 60601-1 Type B and BF, IEC 60601-1-2, according to the
class IIa of the EU medical directive 93/42/EEC