Features:
* Charging cradle with PC data transfer
* LiveTymp progress display
* 4 ipsilateral reflex test frequencies
* Fast pass mode reflex test
* Auto pass/refer evaluation
* Fully customisable test protocol
* Internal database
* AmpliSuite software and NOAH integration included.
* User configurable test settings
* 6 language options: English, Spanish, Portuguese, French, German and Italian
Technical Data:
- Tympanometry Measurements
- Probe Tone: 226Hz +/-2% 85dB SPL +/-2 dB over range 0.2ml to 5ml
- Pressure Range: +200daPa to -400daPa +/- 10daPa or +/-10% (whichever is greater)
- Direction of Sweep: Positive to negative
- Volumetric Range: 0.2ml to 5ml +/-0.1ml or +/-5% (whichever is greater)
- Analysis Performed: Compliance peak level, compliance peak pressure level, gradient and equivalent ear canal volume
- Reflex Measurements
- Reflex Frequencies:: 500Hz, 1kHz, 2kHz, 4kHz +/-2%
- Reflex level (Multi Mode): 85 to 100 dBHL (5 or 10 dB steps)
- Measurement range: 0.01ml to 0.5ml +/-0.01m
- Analysis performed: Reflex maximum amplitude and pass/fail at each test level
- Data Management
- Internal database: 30 patient records
- Optional printer: Designated thermal printer
- Data Transfer: Via USB to ampliSuite, NOAH4 and other EMR systems
- Languages: English, German, Italian, Spanish,French, Portuguese
- sical Data
- Power (Battery Pack): NiMH rechargeable battery
- Dimensions:
- Length: 210mm
- Width: 80mm
- Height: 40mm
- Weight: 650g with cradle / 380g handset
- Standards
- Safety: EN60601-1 / EN60601-1-2
- Impedance: EN60645-5 Type 2 tympanometer
- CE Mark: Complies to EU Medical Device Directive