Coulometric Sensor Method Oxygen Permeability Analyzer Y310 2.0
Introduction
This product is designed and manufactured based on the coulometric sensor method and conforms to ASTM D3985. This instrument can be used to measure the oxygen transmission rate of barrier materials with high and medium barrier properties with high accuracy and high efficiency. Y310 is applicable to the determination of oxygen permeability of plastic films, sheeting, paper, and other packaging materials used in food, rmaceutical, medical apparatus, consumer products, tovoltaic and electronic industries, etc.
Test Principle
Y310 2.0 Oxygen Permeability Analyzer is designed and manufactured based on the coulometric sensor method. The pre-conditioned specimen is clamped in the test oxygen or air flows on one side of the specimen while a stream of high purity nitrogen flows on the other side. Oxygen molecules permeate through the specimen into the nitrogen side and are carried to the coulometric sensor. The sensor analyzes the oxygen concentration and calculates the oxygen transmission rate.
Standard
ASTM D3985ãASTM F2622ãASTM F1927ãASTM F1307ãISO 15105-2ãDIN 53380-3ãJIS K-7126-B