Medical & ITE safety approvals
2MOPP input to output isolation
Suitable for medical equipment up to class BF(∧)
Low Leakage current≤10μA
DOE efficiency level VI
CoC V5 Tier 2(2016)
Standby power ≤0.075W
4.2V to 12V outputs, up to 6W
Interchangeable AC plugs
Protections against: short circuit/over current/ over voltage
Technical Data
Input Voltage:90~264VAC
Input Current:0.2A
Standby Power :0.075W
Output voltage range:±5%
Environmental Requirements
Operationg Temperature: 0~45℃
Storage Temperature:- 20~60℃
Operating Humidity:10~90%
Storage Humidity::5~90%
Operating Altitude:Up to 5,000 M operating altitude
Safety Approvals
Safety aAgency/Mark
Medical
ITE
CB
IEC60601-1
IEC60950-1 IEC62368-1
UL
ANSI/AAMI ES60601-1
UL60950-1 UL62368-1
CAN/CSA C22.2 NO.60601-1
CAN/CSA C22.2 NO.60950-1
TUV-Mark
EN60601-1
-
RCM
-
AS/NZS 60950.1
CCC
-
GB4943.1
PSE
-
J60950-1
KC
-
K60950-1(4.2 to 6VDC model only)
CE
-
EN62368
FCC
-
FCC PART 15
TUV Rheinland-GS
-
EN60950-1 EN62368-1
Mechanical Details
73.5(L)*42.5(W)*30.7(H)mm
Applications:
Medical Device and ITE Equipment