Quick Details
- Processing Time:-
- Port:-
- Supply Ability: -
- Brand Name:-
(PC100E)Handheld ETCO2/Capnometer/Capnograph Monitor with NIBP
Model:PC100E
Parameters:SPO2+NIBP+ETCO2
Features:
1. The Smallest and exquisite handheld vital sign monitor in the world
2. Accurate measurement in 0.05% low perfusion index,the world leading production in the movement environment
3. Provide the parameter of the PI(perfusion index 0.05%-20%)
4. Accurate measurement in all kind of extreme environment measurement such as neonatal and shock patients
5. NIBP adopt SAWAA(Self-Adaptive-Waveform-Amplitude-Adjust)algorithm and SAFD(Self-Adaptive-Fast-Deflation)algorithm
6. Easy operation by 3.5"color TFT screen and Up to 100users ID switch
7. Multiple vital signs measurement parameters for collocation optional
l NIBP,SPO2,TEMP
l SPO2,TEMP
l NIBP,TEMP
l SPO2,ETCO2
l SPO2,NIBP,ETCO2
8. Two measurements mode:Spot,Monitoring
9. Data storage capabilities,support for historical data view and trend chart back to see
l In monitoring mode,can continuously record 48 hours of measurement data
l In spot mode,100users ID,each user can store 200sets data
10. Support audio and visual alarm,support to set alarm uper and lower limit values
11. Lithium ion battery power supply
12. Optional charging dock
Perfomance Specifications:
Display
3.5" color TFT
Resolution
320*480
Indicator
Alarm indicator,Power indicator
Trace
1 plethysmogram waveform
Alarm
Probe off,low power
modes
visual and audio
Application
Neonatal,pediatric and adult patients
Parameters Performance:
NIBP
Adult
PED
NEO
SYS
40~270mmHg
40~200mmHg
40~135mmHg
MEAN
20~230mmHg
20~165mmHg
20~105mmHg
DIA
10~210mmHg
10~150mmHg
10~95mmHg
Pulse Rate Range:40~240bpm
NIBP measurement accuracy
±3mmHg
Pulse rate measurement accuracy
≤2%(≥100bpm)or 2beats/min(<100BPM)
Independent software protection pressure
Adult:300mmHg
PED:240mmHg
NEO:150mmHg
Independent hardware protection pressure
Adult:320-330mmHg
PED:265-275mmHg
NEO:160-165mmHg
Oxygen Saturation(%SPO2)
Measurement Range
Measure Range
SPO2:0-100%
PR:0-254BPM
Perfusion index:0.05%-20%
Accuracy Range
SPO2 saturation:70-100%
Pulse Rate:20-254BPM
Perfusion Index:0.025-20%
Measurement Accuracy
SPO2
Adult(including pediatric):±2digit(70-100%) not defined(<70%)
Neonate:±3digits not defined(<70%)
On motion condition :±3digits
Pulse Rate
Adult(including pediatric):±3digits
Neonate:±3digits
On motion condition::±3digits
Perfusion Index:0.05%-20%
Resolution
SPO2 saturation:1%
Pulse Rate:1BPM
CO2
Measurement range:
ETCO2:0-150mmHg(0-19.7%)(0-20kPa) RR:3-150BPM
Accuracy range:
ETCO2 Range
Accuracy
0-40mmHg
±2mmHg
41-70mmHg
±5% of reading
71-100mmHg
±8% of reading
101-150mmHg
±10% of reading
RR Accuracy:±1BPM
Payment
Shipping