TECHNICAL SPECIFICATIONS
Normal Conditions of Use
Display Resolution
0.1°C (0.1°F)
Operating Temperature
10 to 40°C (50 to 104°F)
Storage Temperature
0 to 50°C (32 to 122°F)
Humidity Rate
<85%
Power
DC 3V (2 x “AA” batteries)
Size
128 x
74x 36 mm / 5 x 2.9x 1.4 in (L x W x H)
Weight
Gross 125.4g / Net 104.5g
Measuring Range
In Body Mode
32.0 to 42.5°C (90 to 108°F)
ln Surface Temp Mod
0 to 60°C (32 to 140°F)
Accuracy
±0.3°C (0.54°F)
Measuring Distance
5 cm – 15 cm (2 in – 5.9 in)
Automatic Stop
7 sec.
Non-contact Body Infrared Thermometer Precision
32 to 35.9°C / 93.2 to 96.6°F
±0.3°C / 0.5°F
According to ASTM Standard E1965-1998 (2003)
36 to 39°C / 96.8 to 102.2°F
±0.2°C / 0.4°F
39 to 42.5°C / 102.2 to
±0.3°C / 0.5°F
108.5°F
ADVICE FOR THE 1ST USE
For stable and reliable results, it is essential to check the Non-contact Body infrared Thermometer and change as needed, as follows:
1st step: take the temperature of a person using a conventional thermometer; you will get 37.5°C (99.5°F) for instance.
2nd step: take the temperature of the same person using the Non- contact Body infrared Thermometer keeping the 5 to 15 cm (2 in to 5.9 in) distance between the thermometer
and the forehead (Take care to remove any obstacle which could alter the measurement (hair,
perspiration…). lf you get 37.5°C (99.5°F), the Non-contact Body infrared Thermometer is properly set and ready for
use.
lf you get a lower temperature, 36.4°C (97.4°F) for example, your
difference is 1.1°C (2.2°F). You should adjust
the temperature on the Non-contact
Body infrared Thermometer and add the difference, i.e.
1.1°C (2.2°F).
To do it, press the “MODE “button for 2 seconds,
the screen displays F1, press “MODE“ button again until you get F3. Press “UP“ button in order to add the difference (in our example, 1.1°C –
2.2°F).
3rd step: To check, take the temperature again using the Non-contact Body infrared Thermometer.