Quick Details
- Processing Time:-
- Port:-
- Supply Ability: -
- Brand Name:-
Multi Parameter Patient monitor with Vital Signs ECG SPO2 NIBP Temp ambulance patient monitor ICU vital signs monitor
12.1" Portable Multi parameter Patient Monitor general ward and clinic monitoring bedside cardiac monitor Beta 12S
Portable Patient Monitor Quick Details:
Central Station
Touch Screen
IoC/BIS (12.1 inch only)
Suntech NIBP/Massimo SPO2
Large 12.1 “ TFT LCD display
7 Channel selectable waveforms
Programmable NIBP cycles
Programmable alarms
Simultaneous display of all parameters
Hung-Kee Patient monitor Brochure
Portable Patient Monitor Technical details:
Table ECG Specifications
Parameter
Specifications
CMRR (common mode rejection ratio)
>90 dB
HR measurement and alarm range
Adult/Pedi: 15bpm - 300bpm
Neonate: 15bpm - 350bpm
HR accuracy
1bpm or 1%, whichever is greater
ECG channel
Up to 7channels
ST range
-2.0 mV - +2.0 mV
ST accuracy
0.02 mV or 10%, whichever is greater
Pacemaker pulse rejection capability
Meets AAMI EC 13
ECG leads
5-lead
Alarm indication
Visual and audible indicators
QRS indicator
Audible tone and heart symbol flashes
ECG speed
12.5 mm/s, 25.0mm/s, 50.0 mm/s
ECG gain
×0.25,×0.5,×1,×2,×4, Auto
Bandwidth
Monitor mode: 0.5 Hz - 40 Hz
Operation mode: 1.0 Hz - 25 Hz
Diagnostic mode: 0.05 Hz - 100 Hz
Tall T-wave rejection capability
Exceeds ANSI/AAMI EC 13 minimum recommended
1.2 mV T-wave amplitude
Heart rate alarm response time
<12s
Defibrillator protected
Tested with5kV
Recovery time following defibrillation
< 5s
Signal delay
0.5s
Auxiliary current (leads off detection)
Active electrode:< 100nA
Reference electrode: < 900 nA
Heart ratemeter accuracy and response to irregular rhythm
Ventricular bigeminy: 80 bpm±2bpm
Slow alternating ventricular bigeminy: 60 bpm ±2bpm
Rapid alternating ventricular bigeminy: 120 bpm ±2bpm
Bidirectional systoles: 90 bpm±2bpm
Response time of heart rate meter to change in heart rate
HR change from 80 bpm to 120 bpm: ≤ 8 s maximum
HR change from 80 bpm to 40 bpm: ≤ 10 s maximum
Time to alarm for tachycardia
Vent tachycardia (1 mVpp, 206 bpm):
Gain 0.5, range 6.5 to 8.4s, average 7.2s
Gain 1.0, range 6.1 to 6.9s, average 6.5s
Gain 2.0, range 5.9 to 6.7s, average 6.3s
Vent tachycardia (2 mVpp, 195 bpm):
Gain 0.5, range 5.4 to 6.2s, average 5.8s
Gain 1.0, range 5.7 to 6.5s, average 6.1s
Gain 2.0, range 5.3 to 6.1s, average 5.7s
ECG analog output
0.67 Hz - 40 Hz
Time to alarm for high heart rate
≤10s
Time to alarm for low heart rate
≤10s
( for more detailed technical data please refer to above PDF brochure)
Portable Patient Monitor Description:
The Beta 12S portable patient monitor is part of Hung-Kee’s premier patient monitoring solution designed to enhance patient care and support timely and accurate clinical assessment across general
ward, daily checking and clinic environments. This portable patient monitor, power supplied by 100-240V~,50/60Hz, adopts 12.1'' color TFT LCD displaying real-time date and waveform. It can
synchronously display seven-channel waveform and full monitoring parameters equipped with an optional 48mm thermal recorder. Maximum 64 units portable patient monitor could be connected to the
central monitoring system via wire network to form a network monitoring system.
Portable Patient Monitor Standard Accessories:
5-lead ECG cable, SPO2 sensor, BP cuff, Temperature probe, Lithium battery, User manual, Power cord
Portable Patient Monitor Optional Accessories:
Aluminum trolley, wall mount stand
FAQ for Portable Patient Monitor:
1.Q: How long is the warranty time?
A: 24 Months
2. Q: When do you delivery?
A: Usually in 7-30 days depends on the order qty and available of some specific parts.
3. Q: Could you accept OEM order?
A: Yes, OEM and ODM are our main business.
4.Q: Do you work on SKD project?
A: Yes, we have more than 14 years experience on SKD service, full assembling procedure will be provided, and detailed guidance of local assembling work will be offered.
5. Q: Can I order a sample?
A: Sure, delivery sample by global express in charge.