Quick Details
- Processing Time:3 -5 days after order has been confirmed
- Port:Berlin
- Supply Ability: 50000 Pieces Per Month
- Brand Name:Invasive ICUVentilator
Item No:
JQ-1100
Ventilation mode
IPPV, A/C, PCV, SIMV, SPONT/CPAP, PSV, SIGH, MANUAL
Ventilator parameter
Tidal volume (Vt)
0 ~ 2000 ML
Frequency (Freq)
1 /min ~ 100 /min
Oxygen concentration
21 % ~ 100 %
I:E
4 : 1 ~ 1 : 8
PEEP
0 cmH2O ~ 40 cmH2O
Pressure limit
20 cmH2O ~ 100 cmH2O
Pressure triggering sensitivity (Ptr)
-20 cmH2O ~ 20 cmH2O (Based on PEEP)
Flow trigger sensitivity (Ftr)
0.5 cmH2O ~ 30 cmH2O
SIGH
0 (off) 1/100 ~ 5/100
Apnea ventilation
OFF, 5 s ~ 60 s
Monitoring parameter
Frequence(Freq)
0 /min ~ 100 /min
Tidal volume (Vt)
0 mL ~ 2500 mL
MV
0 L/min ~ 99 L/min
Airway pressure
0 cmH2O ~ 100 cmH2O
Dynamic lung compliance monitoring
1 mL/cmH2O ~ 1000 mL/cmH2O
Oxygen concentration
15 % ~ 100 %
Oscillogram
P-T(Pressure-Time)
F-T(Flow-Time)
P-V Loop(Pressure-Volume Loop)
Alarm and protection
The AC power failure alarm
Power failure or no connection
Internal battery backup low voltage alarm
< 11.3 ± 0.3 V
No tidal volume
No tidal volume within 6 s
High minute volume alarm
5 L/min ~ 99 L/min
Low minute volume alarm
1 L/min ~ 30 L/min
High airway pressure alarm
20cmH2O ~ 100cmH2O
Low airway pressure alarm
0cmH2O ~ 20cmH2O
High oxygen concentration alarm
19 % ~ 100 %
Low oxygen concentration alarm
18 % ~ 99 %
Continuous pressure alarm
(PEEP + 1.5 cmH2O) over 16s
Suffocation warning
5 ~ 60 s
Fan error
Show on screen
Oxygen deficit
Show on screen
The maximum limited pressure
<12.5 kPa
Working condition
Gas source
O2, AIR
Pressure
280 kPa ~ 600 kPa
Voltage
220 V ± 22 V
Power frequency
50 Hz ± 1 Hz
Input power
900 VA (With air compressor)
250 VA (Without air compressor